OPERATION
33
To change the Test type on the model 6330, please press the SET key three times until
the display shows:
or
or
Press the Up () or Down () arrow keys to toggle the test selection between AC, DC,
IR, GND, RUN, and LLT modes; then press the EXIT key to store the setting and return
to the selected test mode or press the SET key if the parameters of the selected test need
to be changed.
5. AC and DC dielectric withstand test modes setting (ACW/DCW)
5.1. To set the Output Voltage
Press the SET key until the display shows:
Use the Up () or Down () arrow keys to enter the desired test voltage, then press the
SET key to advance to the next parameter or press the EXIT key to store the setting and
return to the test mode. The maximum voltage which may be entered is 3.50kV AC or
4.00kV DC. Any attempt to set the value outside these limits will cause the instrument to
signal an ERROR by producing two short beeps. Pressing the Down () arrow key when
Voltage = 0.00kV also will cause the instrument to signal an ERROR by producing two
short beeps.
5.2. To set the High Leakage Current Limit
Press the SET key until the display shows:
Use the Up () or Down () arrow keys to enter the leakage current high limit setting,
then press the SET key to advance to the next parameter or press the EXIT key to store
the setting and return to the test mode. The unit of measure is in milliamperes with
20.00mA AC and 5.00mA DC as the maximum setting and with 0.00mA AC and DC as
minimum setting. The High Leakage Current Limit will act as the high “trip point” for
the Hipot test. If the measured leakage current exceeds the value, the test will fail. Any
Mode = AC
or
Mode = DC
Mode = IR
or
Mode = GND
Mode = RUN
or
Mode = LLT
Voltage = X.XXkV
HI-Limit = XX.XXmA