OPERATION
The RESET button is not active in this situation. Only the EXIT key will allow you to
return to the test mode. When the EXIT key is pressed the instrument will continue with
its normal failure indication process. The failure light and alarm can be cleared by
pressing the RESET button. If Output Error occurs please call the Slaughter Company
Customer Support Center at 1-800-504-0055 for assistance.
8. Fatal Error
If the instrument has a Fatal Error failure then the following screen will appear:
None of the buttons and keys will be active in this situation. This type of failure
permanently locks the instrument in the Fatal Error mode and requires that the instrument
be serviced by the Slaughter Company. The customer should contact the Slaughter
Company Customer Support Center at 1-800-504-0055 to receive further instruction.
The following Fatal Error identification numbers will represent the type of the failure that
has occurred:
FATAL ERROR 9001 will appear on the display, if the instrument has a recognizable
internal component failure.
FATAL ERROR 9002 will appear on the display, if the instrument’s System data or the
Model/Option data are corrupted and do not match the setting.
FATAL ERROR 9003 will appear on the display, if the instrument’s Calibration data is
corrupted.
FATAL ERROR 9004 will appear on the display, if the instrument has an internal data
transfer error.
OUTPUT ERROR!
EXIT to continue
FATAL ERROR 900X
Call 18005040055