EasyManua.ls Logo

Slaughter 6330 - Functional Run Test Mode; Settings; Measurement

Slaughter 6330
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SPECIFICATIONS
15
Accuracy: (0.1 + 0.05 sec)
Milliohm Offset
Range: 0 100 m
Resolution: 1 m
Accuracy: (2 of setting + 2 m
FUNCTIONAL RUN TEST MODE
DUT POWER
AC Voltage
0 - 277.0 V, Single phase unbalanced, 0 - 30.0 A maximum
Current
30 A maximum continuous
Power Rating
8400 W maximum
DUT Protection
Short Circuit current 50 A < 3 s, Inrush Current 180 A Response
time 10 us.
SETTINGS
Range
Resolution
Accuracy
HI and LO Limit AC
Voltage, V
0.0 - 277.0
0.1
± (1.5% of setting + 0.2
V), 60.0-277 V
HI and LO Limit AC
Current, A
0.0 - 30.0
0.1
± (2% of setting + 2
counts)
HI and LO Limit AC Power,
W
0 - 8400
1
± (5% of setting + 6
counts)
HI and LO Limit Power
Factor
0.000 - 1.000
0.001
W/VA, calculated and
displayed to three
significant digits
HI and LO Limit Leakage
Current
0.00 - 10.00, Hi-
Limit 0=OFF
0.01
± (2% of setting + 2
counts)
Delay Time, second
0.2 - 999.9
0.1
± (0.1% + 0.05 sec)
Dwell Time, second
0, 0.1 - 999.9
(0=continuous)
MEASUREMENT
Range
Resolution
Accuracy
Voltage, Vac
0.0 - 277.0
0.1
± (1.5% of reading + 2
counts), 60.0~277 V
Current, Aac
0.0 - 30.0
0.1
± (2% of reading + 2
counts)
Power, Watts
0 - 8400
1
± (5% of reading + 6
counts)
Power, Factor
0.000 - 1.000
0.001
W/VA, calculated and
displayed to three
significant digits
Leakage Current, mA
0.00 - 10.00
0.01
± (2% of reading + 2
counts)
MD Circuit
Leakage Current measuring resistance = 2K Ω±1%
Timer, second
0.0 - 999.9
0.1
± (0.1% of reading +
0.05 sec)
LINE LEAKAGE
DUT
DUT Input Power
0 - 277.0 V, Single phase unbalanced, 30 A maximum

Table of Contents