OPERATION
50
ACW Pass XXX.Xs
XX-XX X.XXkV XX.XXmA
or
DCW Pass XXX.Xs
XX-XX X.XXkV XX.XXmA
The instrument is ready to perform another test.
1.3. If the operator elects to abort a test in process this can be accomplished by pressing
the RESET button at anytime. The instrument will stop the test process immediately and
the display will show:
ACW Abort XXX.Xs
XX-XX X.XXkV XX.XXmA
or
DCW Abort XXX.Xs
XX-XX X.XXkV XX.XXmA
Please press the TEST button to initiate another test.
1.4. If the leakage current exceeds the high limit setting, but does not exceed the metering
range, the red RESET button will illuminate and an alarm will activate. The display will
show:
ACW HI-Limit XXX.Xs
XX-XX X.XXkV XX.XXmA
or
DCW HI-Limit XXX.Xs
XX-XX X.XXkV XX.XXmA
To stop the alarm, please press the RESET button once. The alarm will stop and the
display will retain the failure information. The instrument is now ready for the next test.
If the RESET button is pressed again, the failure information will be cleared and the
display will indicate the setting data of the test.
1.5. If the leakage current does not exceed the low limit setting, the red RESET button
will illuminate and an alarm will activate. The display will indicate:
ACW LO-Limit XXX.Xs
XX-XX X.XXkV XX.XXmA
or
DCW LO-Limit XXX.Xs
XX-XX X.XXkV XX.XXmA
To stop the alarm, please press the RESET button once. The alarm will stop and the
display will retain the failure information. The instrument is now ready for the next test.
If the RESET button is pressed again, the failure information will be cleared and the
display will indicate the setting data of the test.
1.6. If there is a short circuit in the DUT during the test, the red RESET button will
illuminate and an alarm will activate. The display will show:
ACW OFL XXX.Xs
XX-XX X.XXkV >30.0mA
or
DCW OFL XXX.Xs
XX-XX X.XXkV >5.00mA
To stop the alarm, please press the RESET button once. The alarm will stop and the
display will retain the failure information. The instrument is now ready for the next test.
If the RESET button is pressed again, the failure information will be cleared and the
display will indicate the setting data of the test.