OPERATION
The instrument is ready to perform another test.
5.3. If the operator elects to abort a test in process this can be accomplished by pressing
the RESET button at anytime. The instrument will stop the test process immediately and
the display will show:
LLT Abort
XXX.Xs
XX-XX XXX.XV
XXXXuA
Please press the TEST button to initiate another test.
5.4. If the DUT input voltage exceeds the high limit setting or does not exceed the low
limit setting, the red RESET button will illuminate and an alarm will activate. The
display will show:
LLT Volt-HI
XXX.Xs
XX-XX XX.XV
XXXXuA
LLT Volt-LO
XXX.Xs
XX-XX XX.XV XXXXuA
To stop the alarm, please press the RESET button once. The alarm will stop and the
display will retain the failure information. The instrument is now ready for the next test.
If the RESET button is pressed again, the failure information will be cleared and the
display will indicate the setting data of the test.
5.5. If the DUT leakage current exceeds the high limit setting or does not exceed the low
limit setting, the red RESET button will illuminate and an alarm will activate. The
display will show:
LLT Leakage-HI
XXX.Xs
XX-XX XX.XV XXXXuA
LLT Leakage-LO
XXX.Xs
XX-XX XX.XV XXXXuA
To stop the alarm, please press the RESET button once. The alarm will stop and the
display will retain the failure information. The instrument is now ready for the next test.
If the RESET button is pressed again, the failure information will be cleared and the
display will indicate the setting data of the test.