Page 118 (Model J)
f) Set the reverse jet gain to 430
g) Set the PEEP control to zero and increase the CMV offset
on the reverse jet to achieve 0 mbar.
h) Repeat steps c, d and e until the set and measured agree.
i) Verify PEEP pressure setting by setting the following pressures on the CMV
control panel and verifying the displayed pressure in the pressure display bar.
Set PEEP Pressure Acceptable tolerance:
0 mbar +1 mbar
5 mbar +1 mbar
10 mbar +1 mbar
15 mbar +1 mbar
20 mbar +1 mbar