Page 74 (Model J)
16. Maintenance and Overhaul
The ventilator has the following maintenance / overhaul schedule.
12 months Preventative maintenance Kit: N9010/12/G
24 months (10,000hrs) Overhaul Kit: N9010/24/G
36 months Preventative maintenance Kit: N9010/12/G
48 months (20,000hrs) Overhaul Kit: N9010/48/G
16.1 Preventative Maintenance Kit
The N9010/12/G contains the following:
Conical filter Qty 2 Part Nº: N2185/06
Duckbill valve Qty 2 Part Nº: N2185/05
Duckbill washer Qty 2 Part Nº: N0280/10/04
“O” rings Qty 2 Part Nº: N6618
Orifice block “O” rings Qty 2 Part Nº: N2042
Oxygen cell Qty 1 Part Nº: N2191/10
AA Battery 1.5v Qty 2 Part Nº: M0484/03