Page 340 (Model J)
37.7 Notes on General System Behaviour
The VueLink-SLE5000 communication link is automatically established if the
physical link is present, this communication can take about 40 seconds to
establish. If the physical connection between the SLE5000 and the VueLink
monitor is broken, communication between the two will continue on reconnection
but will take about 40 seconds to start again.
Only one alarm is condition is visible on the SPI at a time. As with the SLE5000,
if there are several alarm conditions present then only the highest priority alarm
will be displayed.
When the mute button is pressed on the SLE5000 to silence the audible alarm,
the alarm condition will clear from the SPI display on the VueLink monitor. The
VueLink monitor will only display unmuted alarms.