(Model J) Page 313
Appendix 2. Check list
The ventilator must pass all the test’s.
Pass Fail
Step 1: Complete power fails alarm test.
Audible alarm triggered and cancelled.
Step 2: Setting FiO
2
.
Measured value corresponds to set.
Step 3: Test of Touch Screen Beep
Touch screen beeps.
Step 6: Oxygen alarm test.
High and Low alarms triggered
Step 11: Function and alarm testing.
Test lung inflates
Step 12:Function and alarm testing.
Analyser corresponds to set pressures
Step 13: Function and alarm testing.
Apnea alarm triggered after 20 seconds
Step 14: Function and alarm testing.
Manual breath delivered
Step 15: Function and alarm testing.
Apnoea alarm triggered after 10 seconds
Step 17: Function and alarm testing.
Continuing positive pressure alarm
is triggered after 10 seconds.
Step 21: Function and alarm testing.
A waveform appears in both waveform
windows
Step 23: Function and alarm testing.
Analyser alternates between set pressures
Step 24: High pressure alarm.
Alarm triggered and cancelled