(Model J) Page 39
8.7 M0915 Power Supply Unit
1. Disconnect the main power supply.
2. Remove the four fixing screws.
3. Disconnect the battery monitor cable (A).
4. Disconnect the main loom power cable
(B).
5. Disconnect the battery monitor cable (C).
6. Disconnect the Mains Present
LED cable (D).
7. Lift the power supply to reveal the mains
inlet cable (E).
8. Disconnect the mains inlet cable and
remove the power supply.
9. Assembly is reversal of removal.
Setup: No setup is required for this device.