(Model J) Page 203
24.2 HFO Ventilation
24.2.1 HFO Only
Frequency Range: 3-20Hz. (1Hz resolution.)
I:E Ratio 1:1
Delta Press Range: 4 mbar to 180 mbar for 3.3 to 6 bar inlet gas pressues
4 mbar to 140 mbar 2.8 to 3.2 bar inlet gas pressues
(Resolution 2mbar)
Mean airway range: 0 mbar to 35 mbar maximum (Resolution 1mbar)
FiO
2
: 21% to 100% (Resolution 1%)
24.2.2 HFO+CMV
BPM: 1 to 150 (Resolution 1BPM)
Inspiratory Time: 0.1 to 3.0 seconds (Resolution 0.01 seconds)
Frequency Range: 3-20Hz. (1Hz resolution.)
I:E Ratio Calculated from BPM and Inspiratory time settings.
(11.2:1 to 1:600)
Inspiratory Pressure: 0 mbar to 65 mbar (Resolution 1mbar)
Delta Press Range: 4 mbar to 180 mbar 3.3 to 6 bar inlet gas pressues
4 mbar to 140 mbar 2.8 to 3.2 bar inlet gas pressues
(Resolution 2mbar)
Mean airway range: 0 mbar to 35 mbar maximum (Resolution 1mbar)
FiO
2
: 21% to 100% (Resolution 1%)