Page 202 (Model J)
24.1.4 PSV
Inspiratory Time: 0.1 to 3.0 seconds (Resolution 0.01 seconds)
CPAP Pressure: 0 mbar to 20 mbar (Resolution 1mbar)
Inspiratory Pressure: 0 mbar to 65 mbar (Resolution 1mbar)
Volume Targeting: 2 ml to 200 ml
From 2ml to 6ml the parameter increments in 0.2ml steps (Fine resolution).
From 6ml to 100ml the parameter increments in 1ml steps (Standard resolution).
From 100ml to 200ml the parameter increments in 5ml steps (Coarse resolution).
FiO
2
: 21% to 100% (Resolution 1%)
Backup breaths are pressure limited, time cycled.
24.1.5 SIMV
BPM: 1 to 150 (Resolution 1BPM)
I:E Ratio: Calculated from BPM and Inspiratory time settings.
(11.2:1 to 1:600)
Inspiratory Time: 0.1 to 3.0 seconds (Resolution 0.01 seconds)
PEEP Pressure: 0 mbar to 20 mbar (Resolution 1mbar)
Inspiratory Pressure: 0 mbar to 65 mbar (Resolution 1mbar)
Volume Targeting: 2 ml to 200 ml
From 2ml to 6ml the parameter increments in 0.2ml steps (Fine resolution).
From 6ml to 100ml the parameter increments in 1ml steps (Standard resolution).
From 100ml to 200ml the parameter increments in 5ml steps (Coarse resolution).
FiO
2
: 21% to 100% (Resolution 1%)
Breaths are flow cycled, pressure limited. Backup breaths are time cycled
pressure limited.