24.2.3 Controls (Via touch screen display)
Adjust button: Activates alarm thresholds for modification
Alarm Auto-Track button: Sets selected % 0
2
for testing
Apnea Sup Button: Sends diagnostic pulse
Apnoea setup button: Activates Backup/Apnea settings panel
Arrow buttons: Increments value
Auto set Button: Autoset high, cycle and low alarm thresholds
Backup parameter control: 1 to 150 bpm (Resolution 1 bpm)
Begin psv Button: Sends diagnostic pulse
BPM parameter control: 1 to 150 bpm (Resolution 1 bpm)
Calibrate button: Calibrates flow sensor
Calibrate Flow Sensor button: Activates calibrate flow sensor panel
Camera button: Stores a loop to memory
CMV button: Selects CMV mode
Confirm button: Confirms selection
Continue without flow button: Selects pressure monitoring operation
Continuous HFO button: Selects oscillations in both inspiratory and expiratory
phases.
Controller button: Activates controller services panel
CPAP button: Selects CPAP mode
CPAP parameter control: 0 to 20 mbar (Resolution 1 mbar)
Delta P parameter control: 4 to 180 mbar (Resolution 2 mbar)
Default button Returns the waveform display to default layout.
End psv button: Sends diagnostic pulse
Expiratory Only button: Selects oscillations in expiratory phase only
F/P button: Selects flow versus pressure loop display
F/V button: Selects flow versus volume loop display
FiO
2
button: Selects fractional concentration of inspired oxygen
against time display
FiO
2
parameter control: 21% to 100% (Resolution 1 %)
Flow button: Activates flow sensor calibration panel