Page 314 (Model J)
Pass Fail
Step 25: Low Pressure alarm.
Alarm triggered and cancelled
Step 26: Breath Not Detected Alarm.
Alarm triggered and cancelled
Step 27: Leak alarm.
Alarm triggered and cancelled
Step 28: Block alarm.
Alarm triggered and cancelled
Step 29: Mains failure alarm.
Alarm triggered and cancelled
Step 30: Gas supply alarms.
No Air supply Alarm triggered and cancelled
Step 31: Gas supply alarms.
No O
2
supply Alarm triggered
Step 32: Gas supply alarms.
No Gas Alarm triggered and cancelled
Step 33: Flow sensor disconnect alarm.
Alarm triggered and cancelled
Step 34: Cycle fail alarm.
Alarm triggered and cancelled
Step 39: HFO waveform test.
ventilator is oscillating, waveforms
appear and I:E ratio = 1:1
Step 42: CMV functional test.
ventilator is cycling, waveforms
appear and I:E ratio = 1:1
Step 44: Wave shape modified.
Pressure wave modified