Page 140 (Model J)
18.1.10 Breath Not Detected Alarm
Step 26 Set the PEEP to 1 mbar.
Remove the test lung from the flow sensor.
Occlude the flow sensor.
After 20 seconds this should initiate an audible and
visual Breath Not Detected alarm.
Refit the test lung to the flow sensor. The Alarm
should self cancel.
Press the reset button to clear the alarm panel of the alarm notification.
Set the PEEP to 5 mbar.