18.1.16 Functional Test of HFO Mode
Step 38 Set the following parameters in the HFO Only preview mode.
HFO Rate ............10Hz
Mean ...................5mbar
FiO
2
.....................21%
Delta P.................Not available in preview mode
After pressing the confirm button the user will be presented with the HFO
screen.
Set the Delta P to 20mbar
Step 39 Verify that:
the ventilator is oscillating,
the oscillation waveform appears in the waveform windows.