Page 146 (Model J)
Note: If functional testing the
ventilator without a flow
sensor fitted the Flow v Time &
Tidal Volume v Time windows
will not appear.
Step 40 Press the Mode Select button and select
HFO+CMV from the Mode Select panel, but do not
press the confirm button.
Note: At this point the user can terminate the functional testing. If the user
would like to functional test all the functions of the ventilator they should
press the Mode Select button and select CMV from the Mode
Select panel, do not press the confirm button. Advance to Step 41.