EasyManua.ls Logo

SLE SLE5000 Model J - Page 131

SLE SLE5000 Model J
344 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
(Model J) Page 131
Depress and hold down the mains power switch for
2 seconds. The ventilator will power down but the
“Complete Power Fail Alarm” will be active.
To cancel the alarm depress the mains switch again
but do not hold down.
Re-connect mains power.
Turn ON the ventilator and wait for it to enter the
Ventilation OFF mode.
18.1.2 Setting FiO
2
Step 2 In the Ventilation OFF mode select the FiO
2
parameter and using
the arrow keys set the value to 21% FiO
2.
The ventilator defaults to
21% on start up.The measured value will be displayed once the O
2
cell has finished it’s self calibration routine. When finished, the set
value and measured value should correspond.
18.1.3 Back-up alarm speaker test
Step 3 Verify that a beep is sounded every time a parameter control is pressed (FiO
2
).
18.1.4 Functional Testing with No Flow Sensor
Step 4 If the ventilator is to be used without a flow sensor
fitted then press the “continue without flow”
button in the alarm panel.
Fit a test lung to the ET manifold.
The user should now advance to section 18.1.6
Note: the functional testing routines have been written with the flow sensor
fitted. The user will be told to skip a section if it applies to testing a flow
related function.

Table of Contents

Related product manuals