Page 158 (Model J)
18.1.20 Functional Test of HFO+CMV Mode
Step 69 For the functional test set the following parameters in the HFO+CMV preview
mode
BPM .....................30
Ti..........................1 sec.
HFO Rate.............10Hz
PEEP....................4mbar
PIP .......................30 mbar
Delta P.................Not available in preview mode
FiO
2
......................21%
After pressing the confirm button the user will be presented with the HFO
+ CMV screen.
Set a Delta P of 4mbar