Page 160 (Model J)
18.1.21 Pressure Change Alarm
Step 73 Disconnect the flow sensor and test lung from
the ET manifold for a maximum of 3 seconds
then reconnect.
Verify that:
the visual and audible Pressure
Change Detected alarm is triggered.
Press the reset button to clear the alarm panel of the alarm notification.
Step 74 Press the Mode Select button and
select Ventilation OFF from the Mode Select
panel.
Press the confirm button to enter the Ventilation
OFF mode.
Complete functional testing is complete.