(Model J) Page 175
The second through the
engineering mode software from the
O
2
& Flow System Calibration
screen.
Warning: The user must not carry out the two point calibration whilst
connected to a patient. The Calibration process only functions in
Ventilation Off mode and would deliver 21% O
2
to the patient for 7 minutes.
Pressing the Two Point O2 Calibration button the user will have to wait 7
minutes before carrying out any new operation.
From the user interface the user the FiO
2
measured value will read CALIB in the
FiO
2
parameter control until the routine is complete.