SLE SLE5000 Model J - Calibration Checklist: Gas Flow and Oxygen Cell Checks

SLE SLE5000 Model J
344 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
(Model J) Page 307
Item 5: Fresh gas flow with SV7 energised.
Item 6: With SV7 de-energised, check the fresh gas flow is within range
Item 7: Monitor the flow from the proximal airway pressure port and ensure that it is
within range.
Item 8: With SV7 energised, and the fresh gas limb occluded, ensure that the fresh
gas pressure relief valve relieves within tolerance
Item 9: Check that the flow past the oxygen cell is within range
Item 10: With SV7 de-energised, and the fresh gas limb occluded, ensure that the
fresh gas pressure is within range
Lower limit Actual reading Upper limit
7.75 lpm 8.25 lpm
Lower limit Actual reading Upper limit
1.2 lpm 2 lpm
Lower limit Actual reading Upper limit
0.15 lpm 0.25 lpm
Lower limit Actual reading Upper limit
140 mbar 170 mbar
Lower limit Actual reading Upper limit
3 lpm 5 lpm
Lower limit Actual reading Upper limit
10 mbar 20 mbar