27 Measured Insp.
Volume
0.1ml 0 to 32768 (0 to 3.2768 l)
28 Measured Exp.
Volume
0.1ml 0 to 32768 (0 to 3.2768 l)
29 Measured PEEP mbar -175 to 175 mbar
30 Measured PIP mbar -175 to 175 mbar
31 Measured FiO2 % 0 to 100 %
32 Measured HFO Delta
P
mbar -175 to 175 mbar
33 Measured HFO Mean mbar -175 to 175 mbar
34 Trigger Count breaths/min Number of patient triggers in the
last minute 0 to 255
35 Measured Minute Vol. ml 0 to 32768 ( 0 to 32.768l)
36 Measured Leak % 0 to 100 % leak
37 Measured Resistance 0.1 mbar.sec/l Airway resistance 0 to 32768 (0
to 3276.8 millibar seconds per
litre)
38 Measured Compliance 0.1 ml/mbar Total Compliance 0 to 255 (0 to
25.5 millilitres per millibar
39 Measured C20/C 0.1 This value is a ratio of the
compliance over the last 20% of
the pressure rise compared to the
total compliance. 0 to 255 (0 to
25.5)
40 DCO
2
1 0 to 32000
41 Current Alarm n/a This value indicates the currently
displayed alarm condition (see
table opposite)
Param. Nº Description Units
Details