(Model J) Page 335
15 Rdyn 0 to 999.9 mbar/l/
s
Measured dynamic resistance. Same as
resist. on the display. Only available with a
flow sensor. Set to zero in HFO modes.
16 Leak 0 to 100 % Patient leak. Only available with a flow
sensor.
17 Trig 0 to 255 Trig/
min
This is the same as the Trigger value. It
display the number of breath triggers in
the last minute. It is set to o in HFO
modes.
18 sPEEP 0 to 35 mbar Set PEEP. Same as the set PEEP on the
display. Set to zero in HFO only mode
19 sTV 3 to 200 ml Set tidal volume. Set to zero in HFO
modes.
20 sPIP 0 to 65 mbar Set PIP. Set to zero in HFO only mode.
21 sFIO_2 0 to 100 % Set FiO
2
value.
22 sHFVAm 0 to 180 mbar Set HFO amplitude. Set to zero in non
HFO modes.
23 sMeanP 0 to 35 mbar Set Mean pressure. Set to zero in non
HFO modes and set to the PEEP setting in
combined mode.
24 C20/C 0 to 25.5 (ratio) Measured C20/C. Only available with a
flow sensor. Set to zero in HFO modes.
25 sMode (ident) 1 = OFF
2 = CPAP
3 = CMV
4 = PTV
5 = PSV
6 = SIMV
7 = HFO
8 = Combin
26 sTrgFl 0.2 to 20.0 l/min Set flow trigger threshold. When a flow
sensor is not available then this value
relates to the pressure trigger threshold.
0.2 = min sensitivity, 20.0 = max sensitivity
27 sAADel 5 to 60 sec Set Apnoea Alarm delay time.
28 highP 10 to 120 mbar Set High Pressure alarm threshold.
29 lowP -120 to 110 mbar Set Low Pressure alarm threshold.
30 lowTV 0 to 200 ml Set Low Tidal Volume alarm threshold
31 highMV 1.0 to 18.0 l Set High Minute volume alarm threshold
32 lowMV 0.0 to 18 l Set Low Minute volume alarm threshold.
33-
63
- - - Dummy Requests.