7-14
XPREZZON
AND QUBE
O
PERATIONS MANUAL
A LARMS
Alarm Recording
An alarm recording prints the last 12 or 20 seconds (based on the Duration setting
found on the Monitor Setup Printer tab), or until the alarm ceases or is suspended,
whichever is greater.
An alarm recording for a parameter can be enabled or disabled in the Module
Configuration Manager (MCM). To enable or disable alarm recordings for those
parameters that are not available in the MCM, you will need to access the Clinical
Setup tab. For more information about Clinical Setup, refer to the XPREZZON
and
qube
System Administration Manual.
If alarm recordings are disabled for a parameter, no alarm recording produces on
either the bedside or the network printer if that parameter goes into an alarm
condition. However, the designated alarm tones still generate and the alarm
messages show.
Printing on a network printer is configuration dependent. Refer to the XPREZZON
and qube
System Administration Manual for details.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Spacelabs qube and is the answer not in the manual?

Spacelabs qube Specifications

General IconGeneral
Display TypeLCD
ConnectivityEthernet
Mounting OptionsWall mount, roll stand
Power SourceAC
CompatibilitySpacelabs
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
Humidity Rangenon-condensing

Related product manuals