EasyManua.ls Logo

Spectrasensors OXY5500 - Ethernet Subnet Mask and Examples; Continuous Measurement Configuration

Spectrasensors OXY5500
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Oxygen Analyzer
418 4900002239 rev. G 7-11-17
Ethernet Subnet Mask
Examples
Configuration of a Continuous Measurement
Precondition: Sensor is connected, sensor constants and calibration values
are already set up correctly (OP-9).
The goal for this configuration is a continuous measurement with a 1 minute
interval, pressure sensor and RTD (Pt100) are deactivated. Instead a manual
fixed value will be transmitted. Refer to Table 4–17.
Table 4–16 Definition of register 5685
Start
Register
Number of
Registers
Reg1/
Reg2
Reg3/
Reg4
Reg5/
Reg6
Reg7/
Reg8
Write
Access
5685 8 Int32:
Ethernet
Subnet Mask
Octet 1
Int32:
Ethernet
Subnet
Mask
Octet 2
Int32:
Ethernet
Subnet
Mask
Octet 3
Int32:
Ethernet
Subnet
Mask
Octet 4
Yes
Table 4–17 Configuration for a continuous measurement
Step Description Register(s) Value
1 Stop the measurement, in case it is
running.
5707, 5708 0 (Int32)
2 Set pressure mode to ‘Manual’. 5705 0 (Boolean)
3 Set manual pressure to “1006.23”. 3147, 3148 1006.23 (Float)
4 Set temperature mode to “Manual”. 5611 0 (Boolean)
5 Set manual temperature to
“20.56”.
2409, 2410 20.56 (Float)
6 Set the interval rate to 1 min.
(60s).
3499, 3500 60 (Int32)
7 Activate the interval mode and
start the continuous measurement
immediately.
5707, 5708 5 (Int32 trans-
lates to 00000101
binary)

Table of Contents