EasyManua.ls Logo

Spectrasensors OXY5500 - TCP;IP Settings; Analog Output Settings (Analogues) Options

Spectrasensors OXY5500
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Oxygen Analyzer
344 4900002239 rev. G 7-11-17
All settings will be applied by pressing .
TCP/IP settings
Use this screen to set TCP/IP. Refer to Figure 3–48.
If DHCP is selected, the IP and Subnet Mask will be assigned by
the DHCP server, and therefore are not editable.
If Static is selected, the IP and Subnet Mask have to be entered
manually. Contact your local Network Administrator for assistance if
input data needs to be confirmed.
Port specifies the network port under which the Modbus application
takes place. The default value for most Modbus applications is 502.
The ID, which will be used with the Modbus communication, can be
set to any value between 1 and 32.
All settings will be applied by pressing .
Analog Output Settings (Analogues) Options
From the MAIN MENU, press Analogues to access the 4-20mA Interface
Settings, 4-20mA Values, Concentration Alarm Relay (LS2) and 4-20mA
Calibration.
Figure 3–48 Digitals -TCP/IP settings

Table of Contents