EasyManua.ls Logo

Spectrasensors OXY5500 - Starting Up the Analyzer

Spectrasensors OXY5500
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
OXY5500 Operator’s Manual 3
11
The ANALOGUES menu is divided into four screens: 4-20mA Interface Settings,
4-20mA Values, Concentration Alarm Relay (LS2) and 4-20mA Calibration.
Refer to “Analog Output Settings (Analogues) Options” on page 3-44.
Press the buttons to navigate between input fields.
To enter editing mode:
1. Press to enter editing mode.
a. Change the setting or value (one digit at a time) using the
buttons.
2. Press the again to save editing changes.
To exit editing mode:
1. Press to cancel and exit editing mode.
Starting Up the Analyzer
Before powering up the OXY5500, refer to the system drawings in Appendix A
to confirm the proper power connections to the power supply, temperature
sensor and oxygen sensor.
As soon as the OXY5500 is connected to the power supply, the analyzer begins
running a short self-test sequence. Refer to Figure 3–15.
All changes will be applied after the next measurement period.
Figure 3–15 Initial screen - Self-check

Table of Contents