EasyManua.ls Logo

Spectrasensors OXY5500 - Required Tools

Spectrasensors OXY5500
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Oxygen Analyzer
B8 4900002239 rev. G 7-11-17
Required tools
Flathead screwdriver (standard size and mini)
9/16” open-ended wrench
Adjustable wrench
10” crescent wrench
To remove the pressure sensor:
1. Turn off the power to the analyzer and open the enclosure door using
a standard flathead screwdriver to unlatch the lock.
2. With a 9/16” wrench, loosen the Swagelok nut closest to the
pressure sensor.
3. Using the same wrench, loosen the Swagelok nut on the T-fitting.
Refer to Figure B–6.
Figure B–6 Removing the Swagelok nuts
PRESSURE
SENSOR NUT
T-FITTING NUT

Table of Contents