EasyManua.ls Logo

Spectrasensors OXY5500 - Measurement Settings Menu

Spectrasensors OXY5500
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
OXY5500 Operator’s Manual 3
5
Measurement Settings Menu
The MEASUREMENT SETTINGS (Meas. Settings) window is selected from the
MAIN MENU screen. Refer to Figure 3–6.
General setting changes for measurements are performed in the
MEASUREMENT SETTINGS menu. If the measurement settings are not
changed, the settings from the last measurement will be applied.
1. Select Meas. Settings from the MAIN MENU screen. A message
window displays requiring confirmation to abort the currently
running measurement. Refer to Figure 3–7.
Figure 3–6 Main menu screen - Measurement
Settings selected
Figure 3–7 Message window - Stop measurements
during configuration

Table of Contents