EasyManua.ls Logo

Spectrasensors OXY5500 - Calibration Temperature Settings

Spectrasensors OXY5500
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Oxygen Analyzer
320 4900002239 rev. G 7-11-17
Calibration Temperature
Use the following instructions to program the analyzer for the correct
calibration temperature. Refer to Figure 3–25.
T0: Temperature at the first calibration point.
Select Auto to measure the temperature at the first calibration
point with the RTD probe (Pt100 temperature sensor).
Select Manual if the first calibration point is known and remains
constant through the calibration process. Temperature values
can be input in
C, F or K. Switch to the desired temperature
unit and change the temperature value in the input field.
T2nd: Temperature at the second calibration point.
Select Auto at the first calibration point for automatic
temperature measurement.
Select Manual to insert the changes to the calibration
temperature manually.
To proceed with the calibration, press the Next button at the top right of the
screen followed by the .
Figure 3–25 Calibration temperature screen

Table of Contents