4-26
17 364 02 0320
Operation
NOTE
If you only want to export treatment data of a patient - without his personal
data - select «OPTIONS» «DATA EXPORT ANONYMOUS» before exporting.
4.5.13 Deleting patient datasets
Open the patient record to be deleted.
Press .
The dataset is set to editing mode.
Press .
Conrm your input by pressing .
h The patient dataset is deleted.
4.5.14 Resetting the treatment shock counter
Press the RESET button in the COUNTER READINGS eld to set the display
of applied shock waves back to ‘0’.
Fig. 4-30 Counter readings
h The display of the number of pulses and the energy output per treatment is reset.
4.5.15 Touch screen calibration
It may be necessary to calibrate the touch screen if you think the touch screen is not
responding correctly.
Press .
In the open menu list, select the function.
The integrated calibration programme starts. You are now requested several
times to press various marking points on the screen until the calibration is com-
plete.
Press each calibration mark briey as it is displayed.