Operation
Triggering shocks
Once all necessary preparations have been taken, it is possible to start the treatment.
• Make sure that the shock wave counter is at zero and a low energy level has been set.
• Press the trigger button on the handpiece.
– Press the trigger button anew to stop shock wave release.
4.8 Treatment with VACU-ACTOR
CAUTION !
Do not apply the VACU-ACTOR in the face.
Do not apply the VACU-ACTOR longer than 30 minutes.
Do not use damaged suction cups.
Only perform treatment in a dry state.
Blood cupping is not allowed!
Only perform treatment at healthy and intact skin!
Avoid hematoma, pain and tissue damage!
• Connect the suction cup into the coupling of the air hose (see Fig. 3-39).
NOTE
The maximum underpressure of the compressor is -935 mbar.
While treatment pressures between -500 mbar and +200 mbar will be reached
depending on the adjusted intensity.
The vent hole limits the maximum underpressure to -500 mbar.
• In case of an error the vent hole can be opened.
Treatment with VACU-ACTOR can either be activated and deactivated with the
footswitch or optional can be switched on and off with the aid of the display. Using
the display the treatment parameters can be adjusted.