Operation
4.6.4 Standard settings
• Before each treatment, make sure that the number of shocks and the actual energy
value are set to zero.
NOTE
Set the nominal value counter to the required value. The '-' symbol appears if zero
is selected. The device then operates without a nominal value specication.
• Start the F-SW treatment at an energy level of 0.1 mJ/mm
2
and a frequency of 6 Hz.
• Start the C-ACTOR II treatment at an energy level of 0.03 mJ/mm
2
and a frequency
of 6 Hz.
• A total of about 2,000 shocks must generally be applied per therapy session.
4.6.5 Treatment
Safety information
Before using the device, the user must make sure it is functioning safely and in proper
condition.
• Each time after the device has been transported, make sure that all functional
checks have been performed on the device before you start treatment. Read about
this in Chapter 4.6.3 funCtiOnal CheCks.
• Read Chapter 1 general safety infOrMatiOn before beginning treatment.
CAUTION!
If the handpiece is not positioned correctly:
Impairment to health due to ineffective treatment is created!
• Dene the treatment zone and make sure that the handpiece
position always corresponds to the treatment zone.
• Make sure that the treatment is only administered by users who
meet the conditions in Chapter 2.2 preCOnditiOns fOr OperatiOn.
• For safety reasons, using the device for applications other than those specied in
Chapter 2.1.1 indiCatiOns is not permitted!
• All status and error messages signaled during treatment must always be attended
to without delay!