General safety information
1.1.2 Safety during treatment of the patient
General note:
Organs with gas pockets, especially parts of the lungs, must NOT be exposed to shock
waves.
As it passes through tissue, the shock wave’s energy is slightly reduced; this reduction
is signicantly weakened by the bone structure.
Pulses can result in adverse cardiac reactions. The patient must be continuously
observed during the treatment and attention must be paid to patient reaction. The
patient should not be under anaesthesia.
Only perform treatments approved by STORZ MEDICAL AG!
The user is responsible for correctly positioning the handpieces and suction cup and
correctly selecting the treatment zone.
Air bubbles reduce the effectiveness of shock waves. Therefore, air bubbles must
always be removed from the shock wave path.
Risk of transmission of microorganisms! Disinfect the handpiece after each use! For
information about this, see Chapter 6 Cleaning, MaintenanCe, Overhaul.
Treatment with the suction cup must be limited to 30 minutes.
1.2 Warning against damage to equipment and
the device
Any damage to the device resulting from incorrect operation is not covered by the
manufacturer’s warranty.
Electromagnetic compatibility
This device complies with the requirements of the applicable standard on
electromagnetic compatibility.
Nevertheless, portable and mobile HF communications equipment (e.g. mobile
phones), including antennas, can interfere with medical electrical equipment. They
should not be used less than 30 cm from the device - including the cables specied by
the manufacturer.
This device is subjected to special precautions regarding EMC and needs to
be installed according the EMC guidelines in
Chapter 7.4.1 eMC guidelines and
ManufaCturer’s deClaratiOn.
The use of accessories or cables that are not authorised by the manufacturer can
result in increased interference emissions or reduced resistance to interference
emissions by the device:
The DUOLITH
®
SD1 is not allowed to be positioned immediately next to or jointly
with other devices. If the operation near or jointly with other devices is required, the
DUOLITH
®
SD1 must be tested in that particular environment to ensure operation
according to technical specication.