EasyManua.ls Logo

Storz DUOLITH SD1 - Standard Settings; Treatment

Storz DUOLITH SD1
142 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-48
17 364 02 0320
Operation
4.7.4 Standard settings
Before each treatment, make sure that the shock counter is set to zero.
NOTE
Set the nominal value counter to the required value. The '-' symbol appears if zero
is selected. The device then operates without a nominal value specication.
R-SW
Start the R-SW treatment at a pressure of 2 bar
eff
and a frequency of 5 Hz.
V-ACTOR
Start the V-ACTOR treatment at an energy level of 2 bar and a frequency of 20 Hz.
The number of shocks for trigger point shock wave therapy (TrST) is different from
ESWT treatment. Such therapies must only be performed by personnel suitably
qualied and trained in TrST.
4.7.5 Treatment
Safety information
Before using the device, the user must make sure it is functioning safely and in proper
condition.
Each time after the device has been transported, make sure that all functional
checks have been performed on the device before you start treatment. Read about
this in Chapter 4.7.3 funCtiOnal CheCks.
Read Chapter 1 general safety infOrMatiOn before beginning treatment.
CAUTION!
If the handpiece is not positioned correctly:
Impairment to health due to ineffective treatment!
Dene the treatment zone and make sure that the handpiece
position always corresponds to the treatment zone.
Make sure that the treatment is only administered by users who
meet the conditions in Chapter 2.2 preCOnditiOns fOr OperatiOn.
For safety reasons, using the device for applications other than those specied in
Chapter 2.1.1 indiCatiOns is not permitted!

Table of Contents

Related product manuals