6 Installation
24
CyFlow™ Ploidy Analyser | Operating Manual | October 2021
6 Installation
6.1 Positioning the device
Carefully select the place where to position and operate the device, it should not be
moved without disconnecting its components. Comply with the following criteria when
positioning the device:
Place the device on a solid, dry and clean horizontal surface.
Place the device and the Sheath Fluid and Waste bottles on surfaces of
identical height, e.g. the same table, at a distance corresponding to the length
of the sensor cables and tubing.
Avoid smoke, dust, vibrations, direct sunlight and any source of direct heat.
Leave a minimum distance of 15 cm between the back of the device and a wall
to keep the ventilation effective.
Do not block the ventilation grid at the back of the device.
Do not place any objects on the device.
6.2 Installation and uninstallation
Installation, connection, disconnection and uninstallation may only be performed by
authorised service personnel and in conformity with the applicable national rules and
regulations.