CyFlow™ Ploidy Analyser | Operating Manual | October 2021
7 Operation
The chapter operation contains all information necessary to operate your CyFlow™ Ploidy
Analyser in a easy, safe and reliable way. It contains notes and information to simplify the
operation of your CyFlow™ Ploidy Analyser.
Sheath Fluid
Sheath Fluid affects the sample flow through the flow cuvette.
• Replace the Sheath Fluid at least once a week
• To ensure most accurate measurement results replace the Sheath
Fluid daily
• After filling up the Sheath Fluid bottle make sure no air bubbles are
trapped in the Inline filter inside the bottle
7.1.1 Switching on the device
Before switching on the device make sure the Sheath Fluid bottle is filled with 800 ml of
Sheath Fluid and is closed with the blue cap. Make sure the Waste bottle is empty and
closed with the red cap.
Configuration on Start-Up
CyView™ starts automatically with the configuration last used.
Requirements
• The Sheath Fluid bottle is filled and connected to the device.
• The Waste bottle is empty and connected to the device.
Procedure
1. Lift up the Display of the device.
2. Switch on the device using the Power switch on the rear side of the device.
3. Push the On switch on the top left corner of the device.
Wait until the device has finished booting.
4. Start the CyView™ software.