11 Maintenance
42
CyFlow™ Ploidy Analyser | Operating Manual | October 2021
11 Maintenance
Servicing is to be performed by an authorised service engineer. Please contact your local
Sysmex representative for further instructions.
Table 12: Maintenance activities
Activity
Qualification requirement
Regular maintenance
Instructed personnel, Qualified personnel,
Authorised service personnel
Other maintenance
Qualified personnel, Authorised service personnel
Technical service
Authorised service personnel
11.1 Maintenance overview
Procedure
Frequency
Chapter
Activity
Priming
After starting the
device for the first
time for the day
7 Operation
Regular
maintenance
After the Sheath
Fluid bottle was
filled
7 Operation
Regular
maintenance
After the Waste
bottle was emptied
7 Operation
Regular
maintenance
Intermediate
cleaning
If necessary
7 Operation
Other maintenance
Shutdown
Daily
7 Operation
Regular
maintenance
Shutdown cleaning
Shutdown,
Every week
or if necessary
7 Operation
Regular
maintenance
Device cleaning
Every week
or if necessary
9 Maintenance
Regular
maintenance
Inline Filter renewal
Every 3 months
or if troubleshooting
applies
9 Maintenance
Other maintenance
Sheath Fluid
refilling
Every day
or if necessary
9 Maintenance
Other maintenance
Waste disposal
Every day
or if necessary
10 Disposal
Other maintenance
Sheath Fluid
disposal
If necessary
10 Disposal
Other maintenance
Preventive
maintenance
Every 12 months
or if troubleshooting
applies
-
Technical service

Table of Contents