7 Operation
CyFlow™ Ploidy Analyser | Operating Manual | October 2021
27
7.2 Running a Quality Check
The Quality Check (QC) procedure will guide the user through the Quality Check of
CyFlow™ Ploidy Analyser with certain reagents
For a detailed decription of the Quality Check procedure refer to the CyFlowPloidy
Analyser QC Manual.
Table 10: Quality Check material
Lightsource
Item
Content
REF No.
UV-LED
DNA control UV
25 ml
05-7302
Green Laser
DNA control PI
25 ml
05-7303
7.3 Performing a measurement (Sample Port)
NOTICE
Sample Preparation
High particle concentrations might clog the device.
Dilute or filter sample if necessary
Mix sample well before analysis
INFORMATION
Refer to specific measurement procedures in the respective assay and
reagent labelling for detailed descriptions.
Different measure modes are available for data acquisition. For more information, please
refer to chapter 13 Software.
Table 11: Measure Modes
Measure Mode
Description
VC electrodes
The start and stop electrode are used to measure a defined
volume of 200 µl
VC of volume
A preset volume is analyzed
Analyze all
The complete sample is analyzed (default setting)
Cells in Region
The sample is analyzed until a preset number of particles
within an specified region is reached