EasyManua.ls Logo

SYSMEX CyFlow Ploidy Analyser - Logging in; Priming the Device; Figure 6: Selecting Prime Mode

SYSMEX CyFlow Ploidy Analyser
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Operation
26
CyFlow™ Ploidy Analyser | Operating Manual | October 2021
7.1.2 Logging in
Each user is assigned to a user role. Each user role has specific privileges. Some
functions of the software are not available for each user role and availability depends on
the assigned privileges. For more information, please refer to chapter 13.1 User
Management.
Requirements
The login dialogue is displayed.
User access must be set up.
Login and password must be known to log into the software.
Procedure
1. Select your name.
2. Enter your password.
3. Click [Login] or press [Enter].
7.1.3 Priming the device
Prime mode is available as a start-up process for the device. Prime mode includes
cleaning and filling the tubes with Sheath Fluid.
Priming should be performed in the following cases:
As a Start-up procedure
As a troubleshooting procedure (no/bad signals, high background, etc.)
As a maintenance procedure (after the Sheath Fluid bottle has been refilled or
after the Waste bottle has been emptied)
INFORMATION
If a CyFlow™ Robby 6 is connected and should be primed the measure
settings for Settings - Measure Mode must be changed to Autoloader.
See Operating Manual CyFlow Ploidy Analyser with Robby 6 (CyView
1.6 SW) IFU GB EN
Requirements
Sample tube with 2 ml Decontamination Solution
Sample tube with 2 ml Sheath Fluid
“Sample Port” in Settings - Measure Mode - Sample Source is selected.
Figure 6: Selecting Prime Mode
Procedure
1. Open a configuration file and select the [Prime mode].
2. Press [Start].
3. Follow the instructions until priming is completed.
4. Click [Continue].

Table of Contents

Other manuals for SYSMEX CyFlow Ploidy Analyser

Related product manuals