14.4 Fluidic specifications
Completely closed system for Sheath Fluid and sample
volumes
No fluid droplets or aerosols generated nor released from
the device
The pressure of the closed system is below 500 mbar
Synthetic quartz flow cuvette with centric flow channel
(capillary diameter 350 μm × 200 μm) for laminar sample
transport with Sheath Fluid
Computer-controlled precision syringe pump for
contamination-free sample transport.
Built-in vacuum pump for Waste bottle; vacuum pressure
is adjustable (computer controlled)
Continuous up to 1200 μl
Minimal sample volume of 850 μl, an additional dead
volume of 50 μl is recommended for best results
200 μl for electrode-based precision absolute counting
(other counting volumes upon request)
Minimal sample volume of 10 μl
10 μl to 1000 μl for syringe-based precision absolute
counting
Sample flow speed adjustable continuously between 0.1
μl/sec and 19,9 μl/sec)
2 glass bottles (each 1 litre) for Sheath Fluid and waste
Avoids sample droplets and sample cross contamination
(computer-controlled)
The CyFlow™ Ploidy Analyser has a closed system for Sheath Fluid and sample material;
no fluid droplets or aerosols generated nor released from the device.