8 Maintenance
34
CyFlow™ Ploidy Analyser | Operating Manual | October 2021
8.2 Maintenance procedures
8.2.1 Sheath Fluid refilling
NOTICE
Bent Tubing
When screwing or unscrewing a bottle and its respective cap, it is
advantageous to twist the bottle instead of the cap.
The Tubing must stay unbent.
Requirements
max. 900 ml Sheath Fluid (REF No. 04-4007_R)
Procedure
1. Unscrew the Sheath Fluid bottle.
2. Pull the tubing, including the Inline Filter, out of the Sheath Fluid bottle.
3. Secure cap and tubing, including the Inline Filter, on a clean surface in a stable
manner.
4. Fill the bottle with up to 900 ml of Sheath Fluid.
5. Insert the tubing, including the Inline Filter, into the bottle.
6. Move the cap gently up and down several times, while keeping the Inline Filter in
Sheath Fluid to remove air bubbles from the Inline Filter.
7. Screw down the cap by turning the bottle. Do not twist the tubing and/or the
sensor cable.
8.2.2 Inline Filter renewal
The used Inline Filter must be disposed of after use according to local legislation.
Requirements
1 Inline Filter (REF No. 04-004-1000)
Procedure
1. Unscrew the Sheath Fluid bottle.
2. Pull the tubing, including the Inline Filter, out of the Sheath Fluid bottle.
3. Secure cap and tubing, including the Inline Filter, on a clean surface in a stable
manner.
4. Pull the used Inline Filter off the tubing (do not pull the tubing off the cap).
5. Dispose of the used Inline Filter.
6. Attach a new Inline Filter to the tubing.
7. Insert the tubing, including the Inline Filter, into the Sheath Fluid bottle.
8. Move the cap gently up and down several times while keeping the Inline Filter
covered in Sheath Fluid to remove air bubbles from the Inline Filter.
9. Screw down the cap by turning the bottle. Do not twist the tubing and the sensor
cable.