13
1.2 DESCRIPTION OF THE EQUIPMENT
1.2.2 Technical specifications
The equipment generates no electromagnetic disturbance.
Class I equipment, with a Type B attached part.
Normal, non-waterproofed equipment.
Equipment not suitable for use in the presence of anaesthetic mixtures inflammable in air,
oxygen or nitrous oxide.
Equipment for continuous use.
Grounding of current under normal conditions max 250 µA.
Europa America
Power 110/230 V ~ 50/60 Hz 110~Vac 50/60 Hz
Absorbed power 150 VA
Maximum mechanical power output of the equipment 500 W
Fuses 2 x F3A 5 x 20mm
EMI EN 60601-1-2
Safety EN 60601-1
EN 957-1 UL 2601-1
Directives 73/23 CEE
93/68 CEE
Weight 98 kg 216 lb
Recommended speed: 50-60 RPM
Noise level below: 70 dB
Electromagnetic resistance system
Adjustment of resistance with 12 different levels of effort
Number of functions: 5
Maximum weight of the user: 110 kg
106
2_Top_En.P65 30/05/2002, 13.5613