1730–Series Introduction
1–13
Table 1–10: Certifications and Compliances (Cont.)
Category Standards or Description
Pollution Degree 4 Pollution that generates persistent conductivity through conductive
dust, rain, or snow. Typical outdoor locations.
Safety Standards
U.S. Nationally Recognized
Testing Laboratory Listing
UL1244 Standard for electrical and electronic measuring and test equipment.
Canadian Certification CAN/CSA C22.2 No. 231 CSA safety requirements for electrical and electronic measuring and
test equipment.
European Union Compliance Low Voltage Directive 73/23/EEC, amended by 93/69/EEC
EN 61010-1 Safety requirements for electrical equipment for measurement,
control, and laboratory use.
Additional Compliance IEC61010-1 Safety requirements for electrical equipment for measurement,
control, and laboratory use.
Safety Certification Compliance
Temperature, operating +5 to +40_ C
Altitude (maximum operating) 2000 meters
Equipment Type Test and measuring
Safety Class Class 1 (as defined in IEC 1010-1, Annex H) – grounded product
Overvoltage Category Overvoltage Category II (as defined in IEC 1010-1, Annex J)
Pollution Degree Pollution Degree 2 (as defined in IEC 1010-1). Note: Rated for indoor use only.
Table 1–11: Physical Characteristics
Characteristic Requirement
Dimensions Height: 5 1/4 inches (133.4 millimeters)
Width: 8 1/2 inches (215.9 millimeters)
Depth: 18 1/8 inches (460.4 millimeters)
Weight Net: 8.5 pounds (3.8 kilograms)