EasyManua.ls Logo

Tektronix 1730 Series - Figure 3-11: Considerations for Custom Installation of an Instrument

Tektronix 1730 Series
258 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1730–Series Installation
3–13
Requires four 0.156” holes below
the 1700F00 cabinet to secure the
instrument to the shelf.
For Flush Front Panel: Cut hole
the same size as the monitor front
molding to allow the monitor front
panel to align with the custom
panel surface.
For Protruding Front Molding:
Cut hole in panel the same size as the
opening in the monitor cabinet to al-
low the front–panel molding to cover
the hole.
Figure 3-11: Considerations for custom installation of an instrument.

Table of Contents

Related product manuals