EasyManua.ls Logo

Tektronix 1730 Series - Figure 5-8: Segment of the Main Board, Showing the Test Points and Adjustment Locations for the +11.8 V Supplies

Tektronix 1730 Series
258 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1730–Series Checks and Adjustments
5–31
A3 MAIN BD
FRONT
U164
R167 R168
U172
–11.8 V
TEST POINT
+11.8 V
TEST POINT
Figure 5-8: Segment of the Main board, showing the test points and
adjustment locations for the +
11.8 V supplies.
b. ADJUST – R167 (–11.8 V ADJ) for –11.78 to –11.82 volts.
c. Connect the voltmeter active lead to the +11.8 V test point. See Fig.
5-8.
d. ADJUST – R168 (+11.8 V ADJ) for +11.78 to +11.82 volts.
6. Adjust 2Line and 1 ms Sweep Calibration
a. Display the CAL signal on the waveform monitor in the 2LINE SWEEP.
b. ADJUST – R660 (Sweep Length) for one cycle of the CAL signal per
major division over the center 10 divisions.
c. Turn on the MAG.
d. ADJUST – R552 (1 ms Cal) for one full cycle over the 10 major
divisions.
7. Adjust 0.2 ms Sweep Calibration
a. Set REF to INT and select CH–A INPUT.

Table of Contents

Related product manuals