Performance Tests
Table
C-3
Settings for Timing Accuracy Checks
SEC/DIV
Time-Mark Generator Setting
Switch
Setting
Xl
(Normal)
XlO
Magnify
X50
Magnify
0.05 µs 50
ns
5
ns
=1
ns
0.1
µs
0.1 µs
10
ns
=2
ns
0.2
µs
0.2 µs 20
ns
=4
ns
0.5
µs
0.5 µs 50
ns
10
ns
1
µs
1 µs
0.1
µs
20
ns
2
µs
2
µs
0.2 µs
0.04µs
5 µs 5 µs 0.5 µs
0.1
µs
10
µs
10
µs
1
µs
0.2 µs
20
µs
20 µs
2 µs
0.4 µs
50 µs 50 µs
5
µs 1 µs
0.1
ms
0.1
ms
10
µs 2 µs
0.2
ms
0.2
ms
20 µs 4 µs
0.5
ms
0.5
ms
50 µs
10
µs
1
ms
1
ms
0.1
ms
20
µs
2
ms
2
ms
0.2
ms
0.04ms
5
ms
5
ms
0.5
ms
0.1
ms
10
ms
10
ms
1
ms
0.2
ms
20
ms
20
ms
2
ms
0.4
ms
50
ms
50
ms
5
ms
1
ms
0.1
s 0.1 s
10
ms
2
ms
0.2
s 0.2 s
20
ms
4
ms
0.5
s 0.5 s
50
ms
10
ms