EasyManua.ls Logo

Tektronix 2212 - Check 500 Hz Trigger Sensitivity

Tektronix 2212
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests
C-20
Procedure Steps
Step
1.
Check
500 Hz Trigger
Sensitivity
a.
Connect
the
low-frequency
sine-wave
generator
output
via
a 50 Q BNC coaxial
cable
and
a 50 Q
termi-
nation
to
the
CH
1 OR X
input
connector.
b.
Set
the
low-frequency
sine-wave
generator
to
pro-
duce
a
3.5-division
display
at
an
output
frequency
of
500 Hz.
c. Set
the
CH 1 VOLTS/DIV
switch
to 1 V/DIV.
d. CHECK -
That
a stable
display
can
be
obtained
by
adjusting
the
Trigger
LEVEL
control
for
each
switch
combination
given
in
Table C-4
with
DC, HF
REJ,
and
AC Trigger COUPLING;
and
that
the
display
will
not
trigger
with
NOISE
REJ
or
LF
REJ
Trigger COUPLING.
Ensure
that
the
TRIG'D light comes
on
when
triggered.
e.
Disconnect
the
test
equipment
from
the
instrument
and
set
the
CH 1 VOLTS/DIV
switch
to
.1
V.
Table C-4
Switch
Combinations for Triggering Checks
Trigger MODE
P-P AUTO
P-P AUTO
NORM
NORM
Trigger SLOPE
Positive
Slope
Negative
Slope
Positive
Slope
Negative
Slope