Procedure Steps
Step
1.
Check
X-Axis
Gain
a. SET:
VOLTS/DIV (CH 1
and
CH
2)
10
mV
Vertical MODE X-Y
Performance Tests
b.
Connect
a 50 m V
standard
amplitude
signal
from
the
calibration
generator
via
a 50 Q BNC
coaxial
cable
to
the
CH 1 OR X
input
connector.
c. CHECK
-That
the
display
is
between
4.85
and
5.15
divisions.
d.
Disconnect
the
test
equipment
from
the
instrument.
Step
2.
Check
X-Axis
Bandwidth
a. SET:
VOLTS/DIV (CH 1
and
CH
2)
50 mV
b.
Connect
the
leveled
sine-wave
generator
output
via
a
50 Q BNC
coaxial
cable
and
a 50 Q BNC
termination
to
the
CH
1 OR X
input
connector.
c.
Set
the
generator
to
produce
an
eight-division
horizon-
tal
display
at
an
output
frequency
of
50 kHz.
d.
Increase
the
generator
output
frequency
until
the
X-Axis (horizontal)
deflection
amplitude
is
5.
7 divi-
sions.
e.
CHECK -
That
the
generators
frequency
is 2 MHz
or
more.
2212 User Manual
C-29

Table of Contents